• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS, INC. PRELUDE SNAP¿; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS, INC. PRELUDE SNAP¿; INTRODUCER, CATHETER Back to Search Results
Catalog Number PLS-1008/A
Medical Device Problem Codes Separation Failure (2547); Activation, Positioning or Separation Problem (2906); Separation Problem (4043); Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
The account alleges that they are unable to tear away the splitable sheath at the valve-ring.The provider had to cut and slice the valve-ring by hand in order to remove the sheath from the implanted lead.No patient injury to report.
 
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Additional Manufacturer Narrative
The suspect device was not returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.Nonconformances of this nature are monitored by the operation team.This complaint will be used to trend for similar complaints.
 
Additional Manufacturer Narrative
The suspect device was returned for evaluation.The device was found to separate inappropriately which is attributed to the manufacturing process.A search of the complaint database was performed and one similar complaint for this lot number were found.The device history record was reviewed, and no exception documents were found.Nonconformances of this nature are monitored by the operation team.This complaint will be used to trend for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRELUDE SNAP¿
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
calle, sor juana ines de la cr
roberto de la madrid
tijuana, b.c. 22630
MX  22630
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC.
calle, sor juana ines de la cr
roberto de la madrid
tijuana, b.c. 22630
MX   22630
Manufacturer Contact
adrian cole
1600 merit parkway
south jordan, UT 84095
8012084225
MDR Report Key23349575
Report Number1721504-2025-00563
Device Sequence Number7956732
Product Code DYB
UDI-Device Identifier10884450467999
UDI-Public10884450467999
Combination Product (Y/N)Y
Initial Reporter CountryGM
PMA/510(K) Number
K143255
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPLS-1008/A
Device Lot NumberI2914410
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/08/2025
Supplement Date Received by Manufacturer11/10/2025
01/15/2026
Initial Report FDA Received Date10/21/2025
Supplement Report FDA Received Date11/18/2025
01/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-