ETHICON INC. SFXSPI MCL+ UNI UD 12IN 4-0 SA PS-2 PMP; SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID
|
Back to Search Results |
|
| Catalog Number |
SXMP1B117 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure in (b)(6) 2025 and barbed suture was used.It was reported that the suture was broken when the surgeon attempted to sew the tissue.Changed another one to complete the surgery.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 10/21/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Date sent to the fda: 11/19/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h3, h6.H3 investigational summary: the product was returned to ethicon for evaluation.One empty foil and one labeled winding former with a needle suture piece were received for analysis.Product code sxmp1b117.During the visual inspection of the returned sample, marks were noted at the edge of the swage area.The suture was examined and damage (scraping) near the swage area was observed, and the end was noted to be cut likely by a surgical instrument.As the suture is absorbable and the duration of exposure of the suture in the environment could not be determined, the functional test could not be performed.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|