| Catalog Number |
1857SI22 |
| Medical Device Problem Code |
Inaccurate Flow Rate (1249)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states, visually inspect the product for any imperfections or surface deterioration prior to use.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.To deflate catheter balloon: gently insert a luer slip syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported they placed a three-way catheter at the end of a turp.They used the bardex ic catheter lot number ngkt1854.Number of us are struggling with the flow on these.As is documented by the nurse they had to come to the bedside and change out this catheter.It appears that there was a flow issue related to this catheter.
|
| |
|
Event or Problem Description
|
|
It was reported they placed a three-way catheter at the end of a turp.They used the bardex ic catheter lot number ngkt1854.Number of us are struggling with the flow on these.As is documented by the nurse they had to come to the bedside and change out this catheter.It appears that there was a flow issue related to this catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|