ALLERGAN (COSTA RICA) STYLE 468 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 468-620
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 08/14/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported right side ¿exchange from textured to smooth implants due to the patient's concerns with the product¿, "capsular contracture baker grade iv" and "discomfort and hardness".Device remains implanted.
 
Additional Manufacturer Narrative
The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: capsular contracture baker grade iv.
 
Event or Problem Description
Healthcare professional reported right side ¿exchange from textured to smooth implants due to the patient's concerns with the product¿, "capsular contracture baker grade iv" and "discomfort and hardness".Device was explanted and it is unknown if it will be returned.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: a5, a6, b5, d6b, h6.
 
Additional Manufacturer Narrative
Device evaluation per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification (if applicable).Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ visibility/palpability: unable to observe through visual inspection as it is a physiological phenomenon.¿ anxiety - product/ procedure: unable to observe through visual inspection as it is a physiological phenomenon.¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (crease, broken device, missing and wear abrasion) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Event or Problem Description
Healthcare professional reported right side ¿exchange from textured to smooth implants due to the patient's concerns with the product¿, "capsular contracture baker grade iv" and "discomfort and hardness".Device was explanted and it is unknown if it will be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 468 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
SANTA BARBARA
900 parkway global park zona f
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key23351825
Report Number9617229-2025-18042
Device Sequence Number9131055
Product Code FWM
UDI-Device Identifier10888628001176
UDI-Public10888628001176
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year1997
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number468-620
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/03/2025
Supplement Date Received by Manufacturer11/21/2025
01/08/2026
Initial Report FDA Received Date10/21/2025
Supplement Report FDA Received Date12/09/2025
01/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight78 KG
-
-