• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE AGG. PLUS SHAVER BLADE SMALL JOINT F-SERIES 2.5MM EACH - ORDER MULTIPLES; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE AGG. PLUS SHAVER BLADE SMALL JOINT F-SERIES 2.5MM EACH - ORDER MULTIPLES; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 0375628000
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the outer tube detached from the device after the nurse cleaned the device.The detached part didn¿t fell into patient¿s body.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Additional Manufacturer Narrative
Multiple attempts have been made to retrieve the device for evaluation, and it is yet to be received in-house.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: device breaking.Probable root cause: inadequate window profile.Inadequate welding process.Improper material strength.Inadequate size selected for the application.Material brittleness.Excessive force applied by the user.Incorrect rfid tag.The reported failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that the outer tube detached from the device after the nurse cleaned the device.The detached part didn¿t fell into patient¿s body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGG. PLUS SHAVER BLADE SMALL JOINT F-SERIES 2.5MM EACH - ORDER MULTIPLES
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
arsalan rizvi
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key23352156
Report Number0002936485-2025-00862
Device Sequence Number9066291
Product Code GFA
UDI-Device Identifier37613327061988
UDI-Public37613327061988
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0375628000
Device Lot Number24082CE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/23/2025
Supplement Date Received by Manufacturer12/10/2025
Initial Report FDA Received Date10/21/2025
Supplement Report FDA Received Date01/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-