| Catalog Number |
68HP-280 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformance's noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".This record is for the right side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: deflation: observed delamination of diaphragm valve through microscopic inspection assessed as adhesive failure also observed crack in the valve assessed as cracked valve seat diaphragm valve.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (creases) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: b5, d9, h3, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".This record is for the right side.The device was explanted and returned.
|
| |
|
Search Alerts/Recalls
|