| Model Number |
H74939171050470 |
| Medical Device Problem Codes |
Inflation Problem (1310); Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b - pro code (product code): fge, lit.(b)(6).G4 - premarket / 510(k) #: k141521, k141597.
|
| |
|
Event or Problem Description
|
|
It was reported that a balloon rupture occurred.The 85% stenosed target lesion was located in the moderately tortuous and mildly calcified arteriovenous fistula.A 5.0 x 40, 75cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon ruptured upon second inflation at 22 atmospheres for 40 seconds.The procedure was completed with a different device.There were no patient complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
G2 - report source (other): updated.D2b - pro code (product code): fge, lit.(b)(6).G4 - premarket / 510(k) #: k141521, k141597.Device evaluated by mfr: the mustang device was returned to our post market quality assurance laboratory.A visual examination identified that the balloon was neatly folded, which indicates that the device was not subjected to positive pressure.A leak test was carried out, and the balloon was inflated to its rated burst pressure for 30 seconds using deionized water and digital timer without issue.A vacuum was then applied.The inflation device was verified at 24 atmospheres, before and after use with calibrated pressure gauge.This inflation to rate of burst pressure was repeated three times with no leaks or drop in pressure noted.No issues were identified with balloon inflation or the balloon material.As per specification, the rated burst pressure for this device is 24 atmospheres.No issues were noted with the tip of the device.A visual examination found no issue with the markerbands.A visual and tactile examination found no kinks or damage to the shaft of the device.This concludes the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that a balloon rupture occurred.The 85% stenosed target lesion was located in the moderately tortuous and mildly calcified arteriovenous fistula.A 5.0 x 40, 75cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon ruptured upon second inflation at 22 atmospheres for 40 seconds.The procedure was completed with a different device.There were no patient complications as a result of this event.
|
| |
|