| Model Number |
D020 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the battery of this implantable cardioverter defibrillator (icd) was suspected to be depleting prematurely and a battery depletion code was recorded.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Technical services (ts) consultant recommended replacement.This icd remains in service and will not be returned.No adverse patient effects were reported.
|
| |
|