• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP FORCEPS; FORCEPS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP FORCEPS; FORCEPS, OPHTHALMIC Back to Search Results
Catalog Number 705.47P
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is provided in sections d.9, h.3, h.6 and h.11.The returned instrument was received in its outer blister, covered with the tyvek foil lid shoeing the lot information.The inner blister which protects the instrument during the shipment was not used.The product shows obvious macroscopic signs of damage.Specifically, the tube is significantly deformed, and the forceps arms are bent as well.There are surgery residues.The instrument was visually inspected with the aid of a photomicroscope using various magnifications.The visual inspection confirmed the macroscopically noticeable damage and displayed the bent forceps arms in detail.Due to the extensive damage, the grasping function of the device could not be tested, nor could any other functional tests be performed.However, the point in time when the described damage occurred can no longer be determined conclusively nor can the situation that led to it be reconstructed.The customer¿s complaint is confirmed as the returned device shows significant deformation, but a root cause for the reported issue cannot be determined.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Non-conformance report provided by originator.A non-conformance based review of the lot number was performed and did not reveal any potential contributing factors to the reported complaint.It cannot be determined how or where the damage to the sample occurred.Therefore, the root cause for the customer's reported event could not be determined.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.The complaint data will continue to be monitored for evidence of adverse trending and further action will be taken, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
The customer reported that during vitreoretinal surgery, the ophthalmic forceps remained fully closed and did not open.The surgery was completed on the same day using another forceps.There was no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP FORCEPS
Common Device Name
FORCEPS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23355233
Report Number3003398873-2025-00395
Device Sequence Number21132696
Product Code HNR
UDI-Device Identifier07612717071148
UDI-Public(01)07612717071148(17)271231(11)250121(10)177N2A
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number705.47P
Device Lot Number177N2A
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer10/17/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/01/2025
Supplement Date Received by Manufacturer10/29/2025
Initial Report FDA Received Date10/22/2025
Supplement Report FDA Received Date11/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/21/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1