| Model Number |
774F75 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.The report of "could not be inflated" was confirmed.During visual inspection balloon was found ruptured at the central area.The balloon edges did not appear to match up at the ruptured region.All through lumens were patent without any leakage or occlusion.No other visible damage or abnormality was observed from catheter body and returned syringe.The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.As part of the manufacturing process, the units go through a balloon inflation and visual inspection.Based on the available information there is no evidence that supports or confirms the failure mode is associated to a manufacturing or design defect.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|
| |
|
Event or Problem Description
|
|
As reported, during testing the balloon of this swan-ganz catheter, it could not be inflated.Issue solved by replaced with new unit.There was no allegation of patient injury.The device was received for evaluation and the balloon was found ruptured at the central area; the balloon edges did not appear to match up at the ruptured region.Patient demographics were not provided.
|
| |
|
Search Alerts/Recalls
|