| Model Number |
IIPECA4010 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/15/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The batch documentation is checked and no anomalies are detected.Any manufacturing fault is ruled out.After examination of the implant, it can be verified that the implant does not show signs of misuse or signs of malfunction that could have contributed to the event.However, it is found that the drilling protocol used is different from the recommended one.
|
| |
|
Event or Problem Description
|
|
The clinician indicates that he placed the implant on (b)(6) 2025 and in the same month it was noted that the implant had not osseointegrated.Patient with bone quality ii.In the event the patient experienced: bone loss and pain.
|
| |
|