• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN INFLATABLE PENILE PROSTHESIS FAMILY; PENILE PROSTHESIS, INFLATABLE Back to Search Results
Catalog Number ES8818
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Erosion (1750)
Date of Event 10/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.The device was not returned for evaluation.Without the benefit of analyzing the device, quality cannot comment on the condition of the device.Should additional information prompt us to alter or supplement any information or conclusions contained in this report, a follow-up report will be submitted.
 
Event or Problem Description
According to the available information the device was explanted and replaced due to erosion/crossover.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN INFLATABLE PENILE PROSTHESIS FAMILY
Common Device Name
PENILE PROSTHESIS, INFLATABLE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key23357917
Report Number2125050-2025-01680
Device Sequence Number8965497
Product Code FHW
UDI-Device Identifier05708932487221
UDI-Public05708932487221
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/31/2025
Device Catalogue NumberES8818
Device Lot Number7493955_ES88182400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexUnknown
-
-