| Model Number |
BF-P150 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Additionally, the investigation found: the adhesive on the bending section cover (a-rubber) is detached.Based on the results of the investigation, the root cause of the reported event could not be determined/identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the bronchovideoscope exhibited that the adhesive on the bending section cover (a-rubber) is detached.There were no reports of patient involvement.
|
| |
|