The bipap a40 device was returned to a third-party service center in reference to the voluntary field safety notice/recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no harm or injury reported.During the evaluation of the device at third-party service center, foam particles were found.The device's software was updated.Additionally, tobacco contamination was observed.Prior to identifying the issue described in fsn 2023-cc-src-039, complaints associated with the issue did not meet requirements for vigilance reporting as a reportable incident.As part of the hhe process, a retrospective review of complaints associated with the issue described in fsn 2023-cc-src-039 was necessary to reassess reportability.Upon this further assessment, these complaints are being reported.
|