• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #2; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6021-0230
Medical Device Problem Codes Material Erosion (1214); Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
Date of Event 09/29/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Patient who underwent tha on (b)(6) 2012.Reoperation was required due to deformation of the stem trunnion and dislodged head.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.
 
Event or Problem Description
No new information.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation involving an accolade stem was reported.The event was confirmed via evaluation of the returned devices.Method & results: product evaluation and results: visual inspection & material analysis: damage consistent with the loss of taper lock was observed on the head and stem of the returned devices.The trunnion of the stem is severely worn.Clinician review: a review of the provided medical records by a clinical consultant indicated: "the x-ray shows a dissociated head and femoral stem.The x-ray shows a dissociated femoral head and stem with a severely damaged trunnion with material loss.No information regarding details of revision surgery was provided.Mechanical failure of an accolade i stem with metal head is confirmed.The root cause for this mechanical complication cannot be determined from the documentation provided." product history review: review of the product history records indicate the reported device was manufactured with no reported relevant discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to deformation of the stem trunnion and dislodged head.The event was confirmed by device evaluation and medical review.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #2
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23358859
Report Number0002249697-2025-01158
Device Sequence Number9202833
Product Code LPH
UDI-Device Identifier04546540510679
UDI-Public04546540510679
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2017
Device Catalogue Number6021-0230
Device Lot Number40718902
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/29/2025
Supplement Date Received by Manufacturer01/30/2026
Initial Report FDA Received Date10/22/2025
Supplement Report FDA Received Date02/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/19/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-