| Model Number |
8886803712 |
| Medical Device Problem Codes |
Material Separation (1562); Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, at skin closure, the staple broke, and a staple went inside without being caught.Another device of the same type was used to resolve the issue.No patient complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: e1.Additional information: d9, g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the instrument was received partially applied with staples remaining.The stapler was jammed with staples misaligned in the cartridge.It was reported that during skin closure, a staple went inside without being caught.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur due to incomplete cycling of the instrument during use.It was additionally reported that during skin closure, the staple broke.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the instrument nose is placed gently on the approximated tissue with the directional arrow in the middle of the incision.Squeeze the trigger completely and release.Failure to completely squeeze the trigger may result in incomplete staple formation and insufficient wound approximation.Repeat this procedure, taking care to space the staples evenly, until the incision is completely closed; any malformed staples should be removed and replaced with a new staple.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|