• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CORPORATION, ASHITAKA RUNTHROUGH; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CORPORATION, ASHITAKA RUNTHROUGH; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 25-1011
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Serious Injury
Event or Problem Description
Terumo medical received a user facility medwatch report # (b)(4).The medwatch event description stated that "the date of event is august 2025.The wire fractured in the coronary artery.Attempted removal of wire, approximately 10 mm of wire left in vessel.Patient had no chest pain and hemodynamically stable when procedure was completed.The procedure was a left heart catheterization." additional information was received on 08oct2025: all known information was reported to medsun.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUNTHROUGH
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
TERUMO CORPORATION, ASHITAKA
150 maimaigi-cho
fujinomiya, shizuok 418-0 015
JA  418-0015
MDR Report Key23359033
Report Number2243441-2025-00055
Device Sequence Number8888375
Product Code DQX
UDI-Device Identifier04987350744340
UDI-Public(01)04987350744340(17)271231(10)250121
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/22/2025
Report Date (Section F) 10/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number25-1011
Device Catalogue NumberTW*AS418XA
Device Lot Number250121Y
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/30/2025
Device Age7 MO
Event Location Hospital
Date Report to Manufacturer10/01/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/22/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age46 YR
Patient SexMale
Patient Weight119 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
Date Report Sent to FDA10/22/2025
-
-