| Model Number |
CNA0T3 |
| Medical Device Problem Code |
Failure to Unfold or Unwrap (1669)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A facility representative reported that during an intraocular lens (iol) implantation procedure, the lens did not unfold properly while being inserted into the patient's eye.There was no harm to the patient, and the procedure was successfully completed.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in d.9., h.3., h.6.And h.11.The product was returned for analysis, and the reported complaint could not be observed.Iol was returned outside of the device.The device was received loose in the product carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been fully advanced outside the nozzle.No damage observed.The lens was returned adhered to the outside of the device.Solution is dried on the iol.No damage observed.The product investigation could not identify the root cause for the reported complaint "lens did not unfold properly" as the lens was returned outside of the device.Due to this, we are unable to confirm the lens and the plunger position for advancement and determine a root cause.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|