| Catalog Number |
8065751761 |
| Medical Device Problem Code |
Suction Problem (2170)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4),.
|
| |
|
Event or Problem Description
|
|
A customer reported that in the during cataract procedure at the start of sculpt phase the handpiece was occluded immediately and continued to alarm with no aspiration.The surgeon paused the procedure and replaced with another handpiece.The surgery was continued as planned.No patient impact was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in sections d.4.D.9.H.3.H.6.And h.11.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.A phacoemulsification handpiece (hp) was returned for testing on this investigation.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for ¿complaints reported against this serial number¿ was performed.No similar complaints were reported for the product serial under investigation.The hp was received for testing on this investigation.A visual assessment of the returned sample found cable damage, connector discoloration, discolored red rot and dented irrigation line.A flow rate test was performed on the irrigation and aspiration lines of the handpiece which found the handpiece to meet specifications per product specifications.The returned sample was connected to a calibrated system.The handpiece tuned successfully and completed a five minute burn-in test with the system set at 100% ultrasonic and torsional power.The handpiece was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements which found the handpiece to meet specifications.The handpiece was found to meet specifications.Therefore, the root cause of the reported event is inconclusive.Manufacture will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Similar incident/event data was collected for the associated reported events.Quality assurance will continue to perform periodic monitoring for evidence of adverse trending and take further action as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|