• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; RSP HUMERAL SOCKET INSERT, 36MM +4MM, STANDARD HXE-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; RSP HUMERAL SOCKET INSERT, 36MM +4MM, STANDARD HXE-PLUS Back to Search Results
Catalog Number 509-00-436
Medical Device Problem Codes Unstable (1667); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Limb Fracture (4518); Joint Laxity (4526)
Date of Event 09/26/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation is completed, any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that the patient required revision surgery due to chronic acromion fracture and instability.
 
Additional Manufacturer Narrative
The agent reported "(the patient has a chronic acromion fracture and instability)".The previous surgery and the surgery detailed in this event occurred 2 years and 7 months apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for review.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.First, there was a nonconformance associated with the main part #509-00-436, rsp humeral socket insert, 36mm +4mm, standard hxe-plus which documents that out of 20 parts lot, 1 item was rejected and scrapped during deburr.Second, there was an ncmr#64800 associated with the concomitant part #508-36-101, glenoid, head w/retaining screw, rsp, 36mm, neutral which document that out of 15 parts lot, 4 items were rejected due to tool marks on polished articulating surface.Later, the rejected were reworked and accepted after proper justification.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was a revision surgery due to acromion fracture and instability.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, prolonged overhead activities, inadequate soft tissue support, patient activities or trauma.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
RSP HUMERAL SOCKET INSERT, 36MM +4MM, STANDARD HXE-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key23366094
Report Number1644408-2025-01573
Device Sequence Number21362898
Product Code KWS
UDI-Device Identifier00888912144582
UDI-Public00888912144582
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K100741
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number509-00-436
Device Lot Number386P1269
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2025
Supplement Date Received by Manufacturer11/26/2025
Initial Report FDA Received Date10/23/2025
Supplement Report FDA Received Date12/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-36-101 LOT: 869C3488
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexMale
-
-