• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRUM AB OPRA IMPLANT SYSTEM; AXOR II

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRUM AB OPRA IMPLANT SYSTEM; AXOR II Back to Search Results
Model Number 1288
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Fall (1848)
Date of Event 08/27/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Sept.26th, 2025: integrum received unit i9130 with a report form stating that the axor ii has detached from the abutment causing the patient to fall multiple times, and that the patient sustained injuries to the eye socket, finger and hips and had to visit the er.Integrum has not received information regarding the specific injuries, i.E.The outcome of the er visit, hence the coding serious injury is a presumtion based on available information.Manufacturing investigation performed, the unit has been manufactured according to specification.Oct.10th, 2025: technical investigation of unit i9130 performed.During the investigation the device functions were tested and components inspected.The reported issue could not be replicated.The device passed the manufacturer's tests.The clamps had some minor marks indicating that the abutment may not have been fully inserted at some point during use, however the root cause of the reported issue could not be established.During service, the clamps were replaced, the device was thoroughly cleaned and functionally tested according to specification with approved results.Integrum will inform the clinic on the importance of donning the axor ii according to the ifu to ensure a stable connection and prevent damage or wear to the clamps.As stated in the ifu, the device shall not be used if a stable connection cannot be achieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPRA IMPLANT SYSTEM
Common Device Name
AXOR II
Manufacturer (Section D)
INTEGRUM AB
gemenskapens gata 9
molndal, 43153
SW  43153
Manufacturer (Section G)
INTEGRUM AB
gemenskapens gata 9
molndal, 43153
SW   43153
Manufacturer Contact
matilda svensson
gemenskapens gata 9
molndal, 43153
SW   43153
MDR Report Key23366232
Report Number3011386779-2025-00246
Device Sequence Number8916169
Product Code PJY
UDI-Device Identifier07340152100634
UDI-Public07340152100634
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P190009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial
Report Date (Section B) 10/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1288
Device Lot Number58767
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2025
Initial Report FDA Received Date10/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-