| Catalog Number |
111654 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Back pain (4895)
|
| Date of Event |
08/27/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, after a thr surgery was performed on (b)(6) 2025, the patient was sent home and post-operatively it was noticed that the prong of one (1) dp bl f/slf rtn ret 63.5 broke off into the patient¿s hip.The patient was informed, and mentioned experiencing back pain, feeling it was due to the retained object.This adverse event was resolved by a removal surgery on (b)(6) 2025 to remove the retained instrument prong.The current health status of the patient remains unknown.
|
| |
|
Search Alerts/Recalls
|