| Model Number |
IPN920760 |
| Medical Device Problem Code |
Incomplete or Inadequate Connection (4037)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that: "the syringe tip does not fit into the puncture needle, making it impossible to check the loss of resistance.This means that the set cannot be used.".
|
| |
|
Event or Problem Description
|
|
It was reported that: "the syringe tip does not fit into the puncture needle, making it impossible to check the loss of resistance.This means that the set cannot be used.".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported complaint of the syringe tip did not fit into the epidural needle was confirmed based on the investigation of the sample received.The customer reported the syringe tip does not fit into the epidural needle.The customer returned one empty kit with one 10ml plastic lor syringe and one epidural needle.Visual examination of the returned sample revealed the epidural needle was an nrfit connection.Based on the product's component list, the kit jc-05400-b typically includes the corresponding epidural needle luer-style connector rz-05400-006 packaged with it.However, it was packaged with an epidural needle nrfit connection instead.A device history record review was performed on the epidural needle and lor syringe with no relevant findings.However, based on the sample received, the potential root cause of this complaint issue is packaging related.Further investigation has been initiated under teleflex quality systems to evaluate this issue.
|
| |
|