| Model Number |
CVPLP2000 |
| Medical Device Problem Codes |
Inappropriate/Inadequate Shock/Stimulation (1574); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Tachycardia (2095); Electric Shock (2554); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
10/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: e7507, e7507 polyhesive ii rem ret el w/cord (lot#unknown); forcefxc, forcefxc force fx generator 110v x1 (serial#(b)(6)) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a c-section, the fellow was using the bovie pencil and was starting to go through the patient's fascia while the surgeon was holding the skin back.The surgeon was on the right side of the patient as well as the grounding pad which was placed on the right thigh, and the generator was also on the right side.The power on the fx was at 35/35.The surgeon said there was a hole in her glove and she was double gloved.The surgeon felt heat and a little jolt.The surgeon felt a pain that went up her right arm and across her chest to her heart.The surgeon reported to feel not right and her heart started to race.The surgeon backed away from the table and they had to call code blue, however, the surgeon was able to ambulate out of the operating room independently, and resuscitation was not required.A new generator and bovie pencil was opened to complete the procedure.The patient and the baby was not harmed/injured.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h1 new information has been received pertaining to the event.This event has been reassessed and the reportability has been determined to be a malfunction.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d4 (expiration date, lot number and udi).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|