| Model Number |
VLFX8GEN |
| Medical Device Problem Codes |
Inaccurate Delivery (2339); Failure to Calibrate (2440); Connection Problem (2900)
|
| Health Effect - Clinical Code |
Numbness (2415)
|
| Date of Event |
09/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the near end of the procedure during skin suturing, the operator used 25w power for coagulation.Large electric sparks appeared on the patient's skin when the pencil was activated, and the pencil leaked electricity causing numbness in the operator's palm.The rem indicator light on the unit remained green and the unit did not issue any alarms.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: e2516h, e2516h disp handswitching pencil x50 (lot#:unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3 additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event does not meet the reporting requirements in 21 cfr 803.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|