• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 IV; FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 IV; FREE THYROXINE Back to Search Results
Model Number FT4 IV
Medical Device Problem Codes Low Test Results (2458); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
1.Describe the event: there was an allegation about questionable results for 1 patient not matching the patient's clinical history when tested for elecsys ft4 iv on a cobas e 801 analytical unit and compared to a different cobas analyzer.2.Patient age range: asku.3.Patient weight range: asku.4.Patient sex breakdown: asku.5.Patient race breakdown: asku.6.Patient ethnicity breakdown: asku.
 
Additional Manufacturer Narrative
1.Summary of investigation results: the investigation is ongoing.2.Summary of follow up/corrective actions taken: the investigation is ongoing.3.Device returned to manufacturer? no.4.Device labeled "for single use?" no.5.Device reprocessed and reused? no.
 
Additional Manufacturer Narrative
1.Summary of investigation results: the investigation did not identify a product problem.The cause of the event could not be determined.2.Summary of follow-up/corrective actions taken: the field service engineer checked the instrument and the sample needle and restored the fluid connection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 IV
Common Device Name
FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key23369426
Report Number1823260-2025-04319
Device Sequence Number9950739
Product Code CEC
UDI-Device Identifier07613336172780
UDI-Public07613336172780
Combination Product (Y/N)N
PMA/510(K) Number
K220456
Exemption NumberVMSR
Number of Events Summarized1
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Expiration Date05/31/2026
Device Model NumberFT4 IV
Device Catalogue Number09043284190
Device Lot Number868732
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer10/31/2025
Initial Report FDA Received Date10/23/2025
Supplement Report FDA Received Date01/16/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-