| Model Number |
LS1037 |
| Medical Device Problem Codes |
Break (1069); Mechanical Problem (1384); Energy Output Problem (1431); Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
07/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted damage to the wires of the device.Functional testing found that without receiving the device, a detailed investigation could not be performed for the reported complaint.It was reported that the hand switch was not reacting to the activation and there was a mechanical failure of the knife when pushing the trigger.The wire is cut and sticking out.The reported issues were confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that, during procedure, the hand switch was not reacting to the activation and there was mechanical failure of the knife (jaw) when pushing the trigger.Photo observation shows that the wire is cut and sticking out.There was no coagulation.The second ligasure after 2 coagulations the jaws clamped the tissue and would not release, after a few clicks the jaws opened and the instrument stopped working.It failed operation and something inside broke.As a result, the patient started bleeding (200 ml) and they had to stop it.To stop the bleeding a clip was placed and the vessel was clamped.Since the patient already had anemia, two doses (500 ml) was administered during the operation as planned.Another ligasure was used successfully with the same generator.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: vlft10gen, vlft10gen ft series energy platformx1 (serial#: unknown), unknown ligasur, unknown ligasure instrument (lot#: unknown) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted there was damage to the wires of the device.Functional testing found that without receiving the device, a detailed investigation could not be performed for the reported complaint.It was reported that the hand switch was not reacting to the activation, and there was a mechanical failure of the jaw.Wire was broken and exposed.The reported issues was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during procedure, the handswitch was not reacting to the activation and there was mechanical failure of the knife (jaw) when pushing the trigger.Photo observation shows that the wire is cut and sticking out.There was no coagulation.The second ligasure after 2 coagulations the jaws clamped the tissue and would not release, after a few clicks the jaws opened and the instrument stopped working.It failed operation and something inside broke.As a result, the patient started bleeding and they had to stop it.Another ligasure was used successfully with the same generator.
|
| |
|