| Model Number |
VTICM5_12.6 (-8.5/2.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
04/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
B5 - describe event or problem: a facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise and lens rotation.The lens was exchanged for a different diopter lens and the problem resolved.Cause of the event was reported as unknown.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise and lens rotation.The lens was exchanged for a different diopter lens and the problem resolved.Cause of the event was reported as unknown.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: the lens was returned in liquid in microcentrifuge vial.Visual inspection found no visible damage to lens.Dimensional and functional inspection found the lens to be within specifications.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.The lens was exchanged for a different diopter lens and the problem resolved.Cause of the event was reported as unknown.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
| |
|
Search Alerts/Recalls
|