| Catalog Number |
20-300 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/17/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reason for reoperation: rupture.A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture".This record is for the right side.The device was explanted and will be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device related to the reported event of rupture was received on november 20, 2025, with lot number 1533100.Per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed broken device through microscopic inspection assessed as fold flaw opening also observed more than three openings assessed as surgical damage and missing piece of shell assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (creases and non penetrating nicks) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture".This record is for the right side.The device was explanted and will be returned.
|
| |
|