• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. EVERA¿ S VR; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. EVERA¿ S VR; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number DVBC3D1
Medical Device Problem Codes Premature Discharge of Battery (1057); Reset Problem (3019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that there was multiple electrical resets and premature battery depletion on the implantable cardioverter defibrillator (icd).It was also noted the patient experienced diaphragmatic stimulation, "jumping or twitching in the chest," and palpitations.The right ventricular (rv) lead remains in use.The device was explanted and replaced. no further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product event summary: performance data collected from the device was received and analyzed.Analysis of the device memory indicated a full electrical reset.Analysis of the device memory showed the battery indicator signifying that it is time for device replacement.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.Analysis of the hybrid revealed the device loss of telemetry and all functions were due to a severely depleted battery.Review of the save to disk data showed that there was an unexplained voltage drop beginning on 11 august 2025.The device was fully functional and operated under normal current drain when powered with an external supply.There was no evidence of a high current drain condition, and no hybrid anomalies were observed.Analysis of the battery revealed plated lithium was found to initiate from the battery case, extend under the head space insulator, and contact the cathode tab.Based on this, the premature battery depletion was determined to be the result of a plated lithium short between the battery case (negative polarity) and the cathode tab (positive polarity).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERA¿ S VR
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23373821
Report Number3004209178-2025-18085
Device Sequence Number10089325
Product Code LWS
UDI-Device Identifier00643169017900
UDI-Public(01)00643169017900(17)170114
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/14/2017
Device Model NumberDVBC3D1
Device Catalogue NumberDVBC3D1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2025
Supplement Date Received by Manufacturer10/27/2025
01/20/2026
Initial Report FDA Received Date10/23/2025
Supplement Report FDA Received Date11/10/2025
02/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
694758 LEAD.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age39 YR
Patient SexFemale
Patient Weight79 KG
-
-