• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 ATTUNE FB TIB BASE SZ 7 POR; KNEE IMPLANT - TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 ATTUNE FB TIB BASE SZ 7 POR; KNEE IMPLANT - TIBIAL TRAY Back to Search Results
Catalog Number 150621007
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Discomfort (2330); Inadequate Osseointegration (2646); Bone Notching (4890)
Date of Event 10/03/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient got a pressfit total knee (b)(6) 2024 and went on a cruise in (b)(6) and was walking 10k steps a day then reported discomfort in the tibia region.Surgeon mentioned that the distal tibial keel showed micromotion and sclerotic bone.Upon removal he noticed the sclerotic bone in that distal tip.Full revision was the result.Doi: (b)(6) 2024.Dor: (b)(6)2025.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB TIB BASE SZ 7 POR
Common Device Name
KNEE IMPLANT - TIBIAL TRAY
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
CORK MFG & MATERIAL WAREHOUSE
loughbeg ringaskiddy
cork
EI  
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23374267
Report Number1818910-2025-18189
Device Sequence Number9950898
Product Code MBH
UDI-Device Identifier10603295507468
UDI-Public(01)10603295507468(17)340531(10)4473890
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K213781
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number150621007
Device Lot Number4473890
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 10/03/2025
Supplement Date Received by Manufacturer10/28/2025
Initial Report FDA Received Date10/23/2025
Supplement Report FDA Received Date10/29/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FB INSRT SZ 8 7MM.; ATTUNE FEM POR CR RT SZ 8.; ATTUNE MEDIAL DOME PAT 41MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-