• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN? X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN? X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) Back to Search Results
Model Number G156
Medical Device Problem Codes Signal Artifact/Noise (1036); Failure to Capture (1081); High impedance (1291); Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range pacing impedance measurements, oversensed noisy signals and intermittent loss of capture (loc) on the right ventricular (rv) channel.Technical services (ts) was consulted and determined and noted it was a non-boston scientific rv lead and suspected the issues were caused by a fracture.No adverse patient effects were reported.This crt-d remains in service.
 
Additional Manufacturer Narrative
Follow up report 1 (additional information): this product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent changes in impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Engineering analysis and testing of returned products has determined that repeated, small movements of the lead terminal ring along with variances in the connection contact force and surface roughness can impact the connection between the spring contact and the lead ring, resulting in intermittent changes in impedance measurement.
 
Event or Problem Description
It was reported that cardiac resynchronization therapy defibrillator (crt-d) exhibited high out of range pacing impedance measurements, oversensed noisy signals and intermittent loss of capture (loc) on the right ventricular (rv) channel.Technical services (ts) was consulted and determined and noted it was a non-boston scientific rv lead and suspected the issues were caused by a fracture.No adverse patient effects were reported.This crt-d remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN? X4 CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23375647
Report Number2124215-2025-76328
Device Sequence Number21074051
Product Code NIK
UDI-Device Identifier00802526534669
UDI-Public(01)00802526534669(17)190321
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/21/2019
Device Model NumberG156
Device Catalogue NumberG156
Device Lot Number375930
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2025
Supplement Date Received by Manufacturer01/28/2026
Initial Report FDA Received Date10/23/2025
Supplement Report FDA Received Date01/29/2026
Date Device Manufactured04/06/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-