• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TM REVERSE 12MM ÿ 130MM LENGTH HUMERAL STEM; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TM REVERSE 12MM ÿ 130MM LENGTH HUMERAL STEM; SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES Back to Search Results
Catalog Number 00434901213
Medical Device Problem Code Material Erosion (1214)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Muscle Weakness/Atrophy (1967); Pain (1994); Loss of Range of Motion (2032); Nonunion/Delayed-union Bone Fracture (2369); Joint Dislocation (2374); Osteolysis (2377); Unequal Limb Length (4534); Bone Notching (4890)
Date of Event 04/03/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial reverse total shoulder on an approximately four (4) years and seven (7) months ago.Subsequently, the patient underwent a revision surgery approximately one (1) year and five (5) years ago due to increased pain and loss of range of motion due to the implant disassociating with clavicular nonunion and scapular malalignment.
 
Additional Manufacturer Narrative
(b)(4).D10: medical products: item#: 00435003600, 36mm vit e liner +0mm; lot#: 64375020.Item#: 115395, comp rvs cntrl 6.5x25mm; lot#: 916180.Item#: 180551, comp lk scr 3.5hex 4.75x20 st; lot#: 428110.Item#: 180553, comp lk scr 3.5hex 4.75x30 st; lot#: 457940.Item#: unknown, unknown tm rev humeral component; lot#: unknown.Item#: 115310, comp rvrs shldr glnsp std 36mm; lot#: 668560.Item#: 010000589, comp rvrs 25mm bsplt ha+adptr; lot#: 919300.G2: foreign: japan.G2: literature: yuki yoshida, takeshi ikegami, jos case reports, https://doi.Org/10.1016/j.Joscr.2025.02.001.H6: component codes: mechanical (g04) - stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: a reverse shoulder arthroplasty was placed in the patient on (b)(6) 2021 at the age of 73.The final routine checkup in 2022 (month unknown) noted, bone resorption of the greater tubercle and scapular notching.At 75 years of age, she began experiencing frequent falls due to muscle weakness eventually requiring admission to a care facility.After admission, there was no documented history of falls or trauma.However, a palpable mass was observed in her right shoulder.Although she experienced pain during shoulder movement while receiving care, the condition was initially monitored without intervention.Her shoulder pain during caregiving gradually worsened, and she was referred to hospital at 77 years of age.On (b)(6) 2024, the patient presented to the emergency department complaining of shoulder pain.Radiography and computed tomography revealed that the palpable mass was the dislocated glenosphere at the anterior surgical site, along with glenosphere dissociation.Revision occurred on (b)(6) 2024 when the glenosphere, liner, baseplate and screws were revised and the humeral stem was retained.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TM REVERSE 12MM ÿ 130MM LENGTH HUMERAL STEM
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION/EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key23378391
Report Number0001822565-2025-03842
Device Sequence Number12967681
Product Code PHX
UDI-Device Identifier00889024268951
UDI-Public(01)00889024268951(17)300713(10)64759372
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K130661
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number00434901213
Device Lot Number64759372
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2025
Initial Report FDA Received Date10/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age77 YR
Patient SexFemale
-
-