|
Additional information: d9, g3, h3, h6.One unpackaged ar-1927bcnf-1, corkscrew®, batch 15434972, was received for investigation.Visual inspection noted that the device was with the sutures and needles attached.The anchor had evidence of damage at the distal end.Functional testing was not performed due to the damage to the device.The method used to prepare the bone, as well as the bone quality encountered during the procedure, was not provided.The most likely cause can be attributed to misuse due to improper bone preparation, misaligned insertion, or prying/leveraging the device during insertion.The complaint allegation is confirmed.
|