| Catalog Number |
8065752212 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A physician reported that an aspiration failure occurred during ultrasound mode and blockage during cataract surgery with intraocular lens implant (iol).Despite the tip and a handpiece were replaced with another ones and reboot was done, the situation could not improve.After replacing the cassette pak with another one, the surgery was restarted.It was unknown whether the surgery was completed or not.There was no patient impact.Additional information has been requested, none received till date.This report pertains to phacoemulsification tip involved in this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in h.6.And h.11.No technical inquiries were received from the customer.One wet gravity fms (fluidics management system) was received at the investigation site; however, no phacoemulsification tip (phaco tip) was returned for evaluation for the report.Therefore, the condition of the product could not be verified.The reported product¿s lot number did not contain a phacoemulsification tip (phaco tip) preventing lot number specific reviews for similar complaints or nonconformances.However, before production release, each product history record is reviewed to ensure that all associated products meet the required specifications and release criteria.Based on the evaluation of the information received, the investigation was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|