| Model Number |
6935M72 |
| Medical Device Problem Codes |
High impedance (1291); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead exhibited t-wave oversensing (twos) on all stored electrograms (egm).Additionally, the rv lead triggered an alert for high rv defibrillation impedance and lead trends have shown chronic high impedance.The lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was further reported that the patient presented to the emergency room due to device was alerting.The device alert was determined to be a result of the rv lead triggered a warning for a spike in rv coil impedance.Additionally, it was noted that the rv lead rv defib impedance trends has increased, a undersensed r wave was noted, r wave variability and tip to ring r wave amplitude variation.It was also reported that the high rate non sustained episodes occurred due rv lead continued to exhibit twos that occurred post sense.
|
| |
|
Search Alerts/Recalls
|