• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LPS CEM FEM STEM 11X150MM BOW; LPS AND S-ROM : KNEE STEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LPS CEM FEM STEM 11X150MM BOW; LPS AND S-ROM : KNEE STEMS Back to Search Results
Catalog Number 198715211
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 09/29/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: h3, h6 investigation summary: no device associated with this report was received for examination.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
It was reported that the patient presented with progressive lucent lines around components and pain, suggesting failed fixation.Implants were removed uneventfully and reimplanted with new components.Frozen section was negative for infection.Cement had not adhered to the bone, but the tibial sleeve was well fixed and thus left in place.The femoral components had been revised on (b)(6) 2024 for failed ingrowth and a cemented stem was placed at that time.Poly was revised (b)(6) 2024 when excessive scar tissue was removed.Affected side: right knee.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LPS CEM FEM STEM 11X150MM BOW
Common Device Name
LPS AND S-ROM : KNEE STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23380101
Report Number1818910-2025-17967
Device Sequence Number8864075
Product Code JWH
UDI-Device Identifier10603295078739
UDI-Public(01)10603295078739(17)320630(10)JR3741
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K003182
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number198715211
Device Lot NumberJR3741
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2025
Initial Report FDA Received Date10/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
Patient SexFemale
-
-