• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEFIBTECH, L.L.C. REVIVER; AUTOMATED EXTERNAL DEFIBRILLATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEFIBTECH, L.L.C. REVIVER; AUTOMATED EXTERNAL DEFIBRILLATORS Back to Search Results
Model Number DDU-C2300
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Although the customer initially reported a service code, review of the aed's internal log files determined that the service code after battery replacement was due to the previous battery fully depleting within the unit.Further analysis indicated that the battery life expectancy was reduced as a result of the customer's failure to promptly address repeated red asi warnings.As a follow up with the customer, the proper maintenance of this device was reviewed.This mdr is being submitted due to the aed becoming non-operational as a result of the battery reaching complete depletion without evidence of user intervention or maintenance following the aeds red asi indicating the unit needs attention or servicing.
 
Event or Problem Description
A customer reported that their aed's asi is flashing red, and reporting a service code.The customer did not report this occurring during patient use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVIVER
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
14 commercial st.
branford CT 06405
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key23381041
Report Number3003521780-2025-01867
Device Sequence Number8888112
Product Code MKJ
UDI-Device Identifier10815098020222
UDI-Public(01)10815098020222
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-C2300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2025
Initial Report FDA Received Date10/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/12/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-