SHIRAKAWA OLYMPUS CO., LTD. VIDEOSCOPE CABLE 150 Back to Search Results
Model Number MAJ-1558
Medical Device Problem Codes Corroded (1131); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was observed that during the device evaluation, the subject device exhibited that flickering/noise/lines and sometimes no image on monitor due to defective scope connector body unit (cable/round connector) and pcb (printed circuit board), external water/chemical seepage inside the pcb resulting into rusting on the components of pcb.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOSCOPE CABLE 150
Common Device Name
VIDEOSCOPE CABLE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23381136
Report Number3002808148-2025-19207
Device Sequence Number18403611
Product Code FDF
UDI-Device Identifier04953170228759
UDI-Public04953170228759
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K051645
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberMAJ-1558
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/26/2025
Initial Report FDA Received Date10/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-