It was reported that a 51-year-old asian female patient underwent a breast reconstruction revision with 700cc mentor memorygel breast implants and experienced bilateral rupture post-operatively, which was diagnosed during surgery.The patient underwent explantation on (b)(6) 2025, and replaced with an 800cc mentor memorygel breast implants.This report is for the right-side device.
|
|
On october 13, 2025, a photo evaluation for the device was completed.Photo evaluation summary: upon visual evaluation of the image provided in the complaint, multiple fragments of the implant were observed wrapped in gauzes.As the product involved in this complaint was discarded, the device could not be analyzed according to our procedures.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|