| Model Number |
0185 |
| Medical Device Problem Codes |
High impedance (1291); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a routine checkup visit a week post generator replacement procedure, this implantable cardioverter defibrillator (icd) device was observed to exhibit high out of range shock impedances on the right ventricular (rv) lead.Technical services (ts) was consulted for further review of the daily threshold and impedance measurements.Ts reviewed the reports and confirmed rv lead shock impedances to have been stable with the old generator, however those rv lead impedances appeared to have increased to be greater than 125 ohms with the new icd device.Ts discussed possible causes and recommended for further evaluation to be performed.At this time, the icd and rv lead remain in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that during a routine checkup visit a week post generator replacement procedure, this implantable cardioverter defibrillator (icd) device was observed to exhibit high out of range shock impedances on the right ventricular (rv) lead.Technical services (ts) was consulted for further review of the daily threshold and impedance measurements.Ts reviewed the reports and confirmed rv lead shock impedances to have been stable with the old generator, however those rv lead impedances appeared to have increased to be greater than 125 ohms with the new icd device.Ts discussed possible causes and recommended for further evaluation to be performed.At this time, the icd and rv lead remain in service.No adverse patient effects were reported.Additional information was received indicated that a follow up visit was performed where xray images were taken.It was suspected that the rv lead to have dislodged at the pin site or possible connection issue however the images were unclear.Technical services (ts) recommended for further evaluation like tug test to be completed.The clinician plans on scheduling patient for an in person visit for further evaluation.At this time, no additional adverse patient effects were reported.This rv lead and icd remain in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product remains implanted and in-service.As such, physical analysis has not been conducted in our laboratory.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This product remains implanted and in-service.As such, physical analysis has not been conducted in our laboratory.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|
Event or Problem Description
|
|
It was reported that during a routine checkup visit a week post generator replacement procedure, this implantable cardioverter defibrillator (icd) device was observed to exhibit high out of range shock impedances on the right ventricular (rv) lead.Technical services (ts) was consulted for further review of the daily threshold and impedance measurements.Ts reviewed the reports and confirmed rv lead shock impedances to have been stable with the old generator.However, those rv lead impedances appeared to have increased to be greater than 125 ohms with the new icd device.Ts discussed possible causes and recommended for further evaluation to be performed.At this time, the icd and rv lead remain in service.No adverse patient effects were reported.Additional information was received indicated that a follow up visit was performed where xray images were taken.It was suspected that the rv lead to have dislodged at the pin site or possible connection issue however the images were unclear.Technical services (ts) recommended for further evaluation like tug test to be completed.The clinician plans on scheduling patient for an in person visit for further evaluation.At this time, no additional adverse patient effects were reported.This rv lead and icd remain in service.Subsequently, additional information was received reported that the physician performed a pocket exploration revision procedure, where it was found that the rv lead was not fully connected to the device.The physician fully connected the rv lead to the device and the impedance issues was successfully resolved.No additional adverse patient effects were reported.The rv lead and icd remain in service.
|
| |
|
Event or Problem Description
|
|
It was reported that during a routine checkup visit a week post generator replacement procedure, this implantable cardioverter defibrillator (icd) device was observed to exhibit high out of range shock impedances on the right ventricular (rv) lead.Technical services (ts) was consulted for further review of the daily threshold and impedance measurements.Ts reviewed the reports and confirmed rv lead shock impedances to have been stable with the old generator.However, those rv lead impedances appeared to have increased to be greater than 125 ohms with the new icd device.Ts discussed possible causes and recommended for further evaluation to be performed.At this time, the icd and rv lead remain in service.No adverse patient effects were reported.Additional information was received indicated that a follow up visit was performed where xray images were taken.It was suspected that the rv lead to have dislodged at the pin site or possible connection issue however the images were unclear.Technical services (ts) recommended for further evaluation like tug test to be completed.The clinician plans on scheduling patient for an in person visit for further evaluation.At this time, no additional adverse patient effects were reported.This rv lead and icd remain in service.Subsequently, additional information was received reported that the physician performed a pocket exploration revision procedure, where it was found that the rv lead was not fully connected to the device.The physician fully connected the rv lead to the device and the impedance issues was successfully resolved.No additional adverse patient effects were reported.The rv lead and icd remain in service.
|
| |
|
Additional Manufacturer Narrative
|
|
Device technical analysis:
this product has not been returned to boston scientific and physical analysis could not be conducted.Device history record review:
a review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Labeling review:
review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Risk assessment:
a risk review was completed and confirmed that the event of shock impedance high out of range was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion:
based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
|
| |
|