| Catalog Number |
3910-400-100 |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient undergoing arthroscopic shoulder surgery experienced a broken drill bit which remained lodged intraosseously.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for investigation.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: drill broke.Probable root cause: application.Excessive force while drilling, selection of wrong drill, reuse of disposable drill, use of reusable drill past lifetime, starting and stopping drill, pulling drill out of bone without running drill, moving drill guide during drilling.The reported failure mode will be monitored for future reoccurrence.
|
| |
|