| Catalog Number |
CL3940 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It is unknown if the device is available for evaluation.The device has been requested to be returned, however, it has not yet been received.
|
| |
|
Event or Problem Description
|
|
The event occurred on an unspecified date and involved a chemolock¿ bag spike with clave¿ additive port that reportedly led to (unspecified) chemotherapy spillage.There was no human harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b3 - date of event - 1 apr 2025.Update: b5 and d9 the reported complaint of a leak could not be confirmed.Without the return of the sample or a photo/video a comprehensive review cannot be performed and a probable cause cannot be determined.The lot history was reviewed; no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|
Event or Problem Description
|
|
The event occurred that involved a chemolock¿ bag spike with clave¿ additive port that reportedly led to (unspecified) chemotherapy spillage during infusion.There was no human harm reported.
|
| |
|