| Catalog Number |
3005S3PX2 |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.2 device(s) was functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.1 device is pending evaluation.Evaluation results findings (components/results): 2 devices: latch / mechanical problem identified, 1 device: appropriate term/code not available / results pending completion of investigation.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between july 1-september 30, 2025.This report summarizes 3 malfunction event(s), where it was reported the device experienced false latch of siderail.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device that was pending evaluation was evaluated in the field and the issue was confirmed.The device was repaired on site and returned to service.Evaluation results findings (components/results) 1 device: siderail/mechanical problem identified.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|