| Catalog Number |
650705550001 |
| Medical Device Problem Code |
Device Fell (4014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.3 device(s) was functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.Evaluation results findings (components/results): 2 devices: gauges/meters / failure to calibrate or used out of calibration, 1 device: hydraulic system / mechanical problem identified.There was no remedial action taken.This device is not labeled for single use.
|
| |
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between july 1-september 30, 2025.This report summarizes 3 malfunction event(s), where it was reported the device experienced cot tip or drop (during load/unload).No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|