|
This complaint was involved with three devices.Device 1 is being reported under mdr 2247858-2025-00254, device 2 is being reported under mdr 2247858-2025-00255, and device 3 is being reported under mdr 2247858-2025-00256.Bolton medical, inc.(d/b/a terumo aortic), herein known as the "company", is submitting this report pursuant to 21 cfr part 803, has made reasonable efforts to obtain complete information, and has provided as much as is available to the company as of the submission date of this report.This report is based on information obtained by the company, which may not have been able to fully investigate or verify prior to the date the report was required by the regulations.Any required fields that are unpopulated are blank because the information is currently unknown, unavailable, or not applicable.This report does not constitute an admission or a conclusion by anyone that the device caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the certain regulations, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
"on (b)(6) 2025, the core lab provided a significant finding of stent fracture regarding subject (b)(6) in review of their 3-year x-ray dated (b)(6) 2025, more specifically a barb break at the proximal aspect of the stent.In comparing the 2-year, the core lab further determined the barb break was already present at 2-year dated (b)(6) 2024.The report for the 2-year x-ray dated (b)(6) 2024 did not specify the barb fracture, however the pi confirmed the initial presence at 2-year and seeing it again on the 3-year imaging.Both the 2-year ct dated (b)(6) 2024 and the 3-year ct dated (b)(6) 2025 reflected a patent device and no evidence of endoleak.The site reported a device deficiency occurring initially for the 2-year timepoint, and no adverse event occurred as a result of the deficiency." patient outcome: "the subject had their 3-year follow-up on (b)(6) 2025 and resulted in the subject having a ct afterwards (details specified above) due to the imaging quality of the dus being limited by bowel gas.Subject is planned to return in 1-year for their next follow-up.".
|