| Catalog Number |
3009PX-LEX-450 |
| Medical Device Problem Code |
Device Fell (4014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Events occurred between july 1-september 30, 2025.This report summarizes 5 malfunction events(s), where it was reported the devices experienced fowler unexpected drop/collapse.No adverse consequence or clinically relevant delay in treatment was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.4 device(s) was functionally/visually inspected in the field.The device(s) was/were repaired and returned to use.1 device was not evaluated and no cause was determined, as the customer did not make the device accessible for testing.Evaluation results findings (components/results).3 devices: shock absorber / wear problem.1 device: appropriate term/code not available / no findings available.1 device: shock absorber / mechanical problem identified.There was no remedial action taken.This device is not labeled for single use.
|
| |
|