• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLAUKOS CORPORATION ISTENT INFINITE TRABECULAR MICRO-BYPASS SYSTEM; AQUEOUS SHUNT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLAUKOS CORPORATION ISTENT INFINITE TRABECULAR MICRO-BYPASS SYSTEM; AQUEOUS SHUNT Back to Search Results
Model Number G2-W-US
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that following the successful implantation of the first stent (of three stents) from a trabecular microbypass stent system in the patient's right eye (od), the trocar broke during attempt to implant the second stent.Per report, the fragments were removed intraoperatively, and a back-up device was used to complete the procedure with no report of patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information: the device is expected to be returned for evaluation but has not been received at glaukos evaluation center.Therefore, product testing on the actual device could not be performed.The device history record review of the manufacturing lot was performed and there were no non-conformities found to be related to the reported event.A complaint history review was completed.The electronic complaint system was searched for the specified lot#.There were no similar issues reported for this lot #.A review of the device labeling including the risk analysis was completed.The reported issue (trocar fragmentation) is identified to be an anticipated hazard.The residual risk for each potential harm was found to be within the insignificant risk region.The ifu adequately provides instructions for stent implantation, precautions, and warnings for the proper use and handling of the device.Mfr# reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISTENT INFINITE TRABECULAR MICRO-BYPASS SYSTEM
Common Device Name
AQUEOUS SHUNT
Manufacturer (Section D)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer (Section G)
GLAUKOS CORPORATION
229 avenida fabricante
san clemente CA 92672
Manufacturer Contact
brittany massey
229 avenida fabricante
san clemente, CA 92672
949367960
MDR Report Key23386138
Report Number2032546-2025-00109
Device Sequence Number18403661
Product Code KYF
UDI-Device Identifier00853704002449
UDI-Public01008537040024491032802017270531
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K220032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG2-W-US
Device Lot Number328020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2025
Initial Report FDA Received Date10/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-