Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D3.Manufacturer email address.G1.Contact office name and email address.G1.Contact office - manufacturer site - email address.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) tsvwh10, (impl tapered scr-v ha 4.7 mm 4.5mm 10mm) for evaluation.Visual evaluation was performed; the implant has signs of use with debris on its external threads.The implants drive feature was damaged.The implant also had a fractured collar, and a fractured screw was stuck inside.The other portion of the screw was not returned.Additional information was obtained on 31-oct-2025 regarding the inspection findings.The customer confirmed that the fractured collar and broken screw occurred during removal of the implant.Dhr review was completed for the subject lot number 1234370.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1234370 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental: functional: damaged drive feature¿ based on the investigation and risk management file review, the most likely root cause determined from the investigation was customer error, applying to much torque/speed during seating.Therefore, based on the available information, a device malfunction did occur.The implants drive feature was damaged.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|