• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIRONA DENTAL SYSTEMS GMBH MIDWEST® PHOENIX¿ PRO PP; HANDPIECE, AIR-POWERED, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIRONA DENTAL SYSTEMS GMBH MIDWEST® PHOENIX¿ PRO PP; HANDPIECE, AIR-POWERED, DENTAL Back to Search Results
Catalog Number 884310
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code Burn(s) (1757)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
1 of 1 devices were returned for evaluation.Evaluation of one device found excessive wear due to a lack of proper maintenance.The device had debris build-up.A friction problem was identified from pressure on the cap.The device did not heat up during evaluation.The device was repaired and returned to the customer.This event is being submitted as part of vmsr.Only one event was received for this device during this quarter.
 
Event or Problem Description
This report summarizes 1 malfunction event where a midwest® phoenix¿ pro pp handpiece overheated.1 event resulted in the patient being slightly burned with no intervention.
 
Additional Manufacturer Narrative
This follow up report provides the requested update to add vmsr to the exemption/variance number field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST® PHOENIX¿ PRO PP
Common Device Name
HANDPIECE, AIR-POWERED, DENTAL
Manufacturer (Section D)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM  D-64625
Manufacturer (Section G)
SIRONA DENTAL SYSTEMS GMBH
fabrikstrasse 31
bensheim, hessen D-646 25
GM   D-64625
Manufacturer Contact
dan eagar
221 w. philadelphia st.
york, PA 17401
7178494593
MDR Report Key23389476
Report Number9614977-2025-00035
Device Sequence Number10034313
Product Code EFB
UDI-Device IdentifierE2768843100
UDI-PublicE2768843100
Combination Product (Y/N)N
PMA/510(K) Number
K131319
Exemption NumberVMSR
Number of Events Summarized1
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number884310
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/27/2025
Supplement Report FDA Received Date02/02/2026
Patient Sequence Number1
Patient SexUnknown
-
-